Autoclaving
1. Eject the pipette tip.
2. Unscrew the pipette shaft from the grip.
3. Autoclave the complete pipette shaft without
any further disassembling.
4. Allow the pipette shaft to completely cool
and dry.
5. Screw the pipette shaft into the grip again.
6. Perform a reference run (rEF).
The pipette shaft of the Transferpette
®
electronic (highlighted in
picture) can be autoclaved at 121 °C (250 °F) at a pressure of 2 bar
(30 psi) with a holding time of at least 15 minutes according to
DIN EN 285.
The handgrip can not be autoclaved!
The effectiveness of the autoclaving
must be verified by the user.
Maximum reliability is obtained with
vacuum sterilization. We recommend
the use of sterilization bags.
Reference run (rEF)
1. Bring up rEF mode
2. Perform the
reference run
Simoultaneously press the
MENU and the ENTER key
to activate the rEF mode.
Press the pipetting key once
to start the reference run. A
noise can be heard,
clearly indicating the function
is being performed.
A manual refernce run must be completed each time the pipette shaft is reattached to
the handle. The reference run is needed to assure secure connection of the piston.
If the pipette shaft is autoclaved frequently, then the
piston should be oiled with the supplied silicone oil in
order to provide smoother movement.
Attention:
Note:
What to do How to do it Keys to press Display readout
60
After the reference run, the display automatically returns to the previous program.